Техническая информация
-
Gas Analyzers In-situ gas analyzer Model Name GM32 The GM32 in-situ gas analyzer measures SO2, NO, NO2 and NH3, as well as TRS components in Kraft pulp processes including pressure and temperature in the gas duct. Direct, fast and without gas sampling and transport. This means control systems are able to work with actual values in real time. Thanks to self-monitoring the measured value are reliable at any time. In case of a malfunction an early warning will be shown. For emission measurement of SO2 and NO, GM32 is tested for suitability according to European standard EN 15267-3. An ATEX version with comprehensive safety functions and sophisticated system technology is also available, as well as the GM32 TRS-PE (PE = Pulp Emission) for TRS measurement in Kraft pulp processes.
At a glance
Direct, fast in-situ measurement No gas sampling, no gas transport, no gas conditioning Up to eight measuring components at the same time, plus process temperature and pressure DOAS and CDE evaluation process Numerous independent measuring ranges with consistent accuracy Automatic self-test function (QAL3) without test gases Overpressure encapsulated design for ATEX Zones 1 and 2 Your benefits
Measured values in real time without altering the gas composition Short-term process deviations are detected Representative measurement byselection of cross-duct or measuring probe versions Fast on-site service due to modular design Long maintenance-free intervals Cost-effective in-situ gas analyzer - also in ATEX design Low cost of installation and operation, no test gases required Complete emissions measurement in Kraft pulp process by way of simultaneous measurement of TRS components plus SO2, NO, NH3 Fields of application
For monitoring and control in flue gas purification,such as denitrification plants, desulfurization plants and converters Emissions monitoring according to EU directives or local regulations Monitoring of landfill gases Monitoring tasks, for example, in nitric acid and ammonia production processes TRS emissions monitoring in the Kraft pulp industry Note
The exact device specifications and performance data of the product may deviate from the information provided here, and depend on the application in which the product is being used and the relevant customer specifications.
Technical data Measurement principles: Chemometrical data evaluation (CDE), Differential optical absorption spectroscopy (DOAS) Measured values: (CH3)2S, (CH3)2S2, CH3SH, H2S, NH3, NO, NO2, SO2, TRS Max. number of measurands: 8 Performance tested measurands: NO, SO2 Measuring ranges: (CH3)2S: 0 ... 5 ppm / 0 ... 18 ppm
(CH3)2S2: 0 ... 6 ppm / 0 ... 12 ppm
CH3SH: 0 ... 7 ppm / 0 ... 23 ppm
H2S: 0 ... 16 ppm / 0 ... 33 ppm
NH3: 0 ... 30 ppm / 0 ... 2,600 ppm
NO: 0 ... 40 ppm / 0 ... 1,900 ppm
NO2: 0 ... 50 ppm / 0 ... 1,000 ppm
SO2: 0 ... 15 ppm / 0 ... 7,000 ppm
TRS: 0 ... 26 ppm / 0 ... 65 ppmComment: For device variant 7: TRS = H2S + CH3SH + (CH3)2S + 2x (CH3)2S2
The measuring ranges depend on application
Measuring ranges refer to 1 m measuring path
TRS = H2S + CH3SH (as H2S equivalent)Length of measuring path: Cross-duct version: 0.4 ... 12 m, Measuring probe version: 0.227 ... 1.75 m
measuring ranges depend on probe typeProcess temperature: 0 ... +550 °C
versions for higher temperatures on requestProcess pressure: 60 ... 200 hPa
relativeAmbient temperature: -20 ... +55 °C
temperature change ±10 °C/h maximumAmbient humidity: 0 ... 96 %
relative humidity; bedewing of optical surfaces not permittedConformities: 2000/76/EC
2001/80/EC
27. BImSchV
EN 14181
EN 15267
GOST
MCERTS
U.S. EPA compliantEx-approvals ATEX: II 2G Ex px [op is Gc] IIC T3 Gb
II 3G Ex pz op is IIC T3 GcElectrical safety: CE Enclosure rating: Ex-version: IP 65
Standard version: IP 65, IP 69KAnalog outputs: 2 outputs:
0/4 ... 22 mA, 500 Ω
per module, modules can be selected and extended as requiredAnalog inputs: 2 inputs:
0/4 ... 22 mA, 100 Ω
per module, modules can be selected and extended as requiredDigital outputs: 4 outputs:
48 V AC/DC, 0.5 A, 25 W
per module, modules can be selected and extended as requiredDigital inputs: 4 inputs:
3.9 V, 4.5 mA, 0.55 W
per module, modules can be selected and extended as requiredInterfaces: Ethernet
extendable via external operating unit (option)Bus protocol: Ethernet TCP/IP
OPCOperation: Via integrated operating unit or SOPAS ET software Device version: Cross-duct version
Ex-version
Measuring probe versionSystem components: Measuring probe or reflector unit
Sender/receiver unit
SCU control unit (option)Corrective functions: Internal correction of contamination Test functions: Internal zero point check
Control cycle for zero and span point according to QAL3
Заказать оригинальный Газоанализаторы GM32 – маркировка изделия бренда SICK и с розничной ценой за штуку. Купите GM32 по специальной цене оптом или в розницу в компании ДАНЭКС. Специальная цена распространяется на продукцию SICK для наших клиентов.
Купить с доставкой GM32 SICK в ДАНЭКС в розницу или оптом можно по телефону, отправить заявку на почту или воспользоваться формой обратной связи на нашем сайте. Доставка по России, Казахстану, Белоруссии и в другие страны СНГ.
Подобрать аналоги для Газоанализаторы SICK или других брендов и уточнить актуальную цену на GM32 вы сможете, написав нам E-mail или позвонив нам по телефону номеру, указанному на каждой странице.
Скачать
- Документация
- Инструкция
- Сертификаты и декларации
- Поиск описания GM32 на сайте проиводителя тут
Доставка
Самовывоз
- FCA г. Санкт-Петербург склад Продавца
Транспортной компанией
- Деловые Линии
- СДЭК
Гарания
- Гарантийный срок на поставляемый Товар составляет 12 месяцев с момента получения Товара Покупателем, если иной срок не установлен в паспортах и сертификатах на Товар.